AJ-HX630
Imaging Diagnostic Equipment
X Ray Equipment
CE, ISO13485
50kw
10-630 Ma
0.003~6.3s (34 Steps) Adjustment
30kHz/400kHz(Optional)
73*200cm, Move Freely, The Wheels with Brake.
ANJUE
Standard Export Packing
ISO; CE
China
Product Description
630mA 50kW High Frequency X Ray Machine AJ-HX630 with Bucky Stand

Techinical Parameters
Model | AJ-HX630 |
Power | 50kW |
Power supply | 50HZ,60HZ, wire diameter >4mm2, internal resistance<0.5Ω; 20/32/50kW(220V /380V available),65kW(only 380V). |
mA | 10-630 mA |
KV | 40-150kV stepping by 1kV |
Exposure time | 0.003~6.3s(34 steps)adjustment:Increased /reduced by 1mS each time |
Focus size | Default 0.6*0.6mm/1.2*1.2mm dual focus, different brands of tubes optional |
Anode rotating speed/Anode heat capacity | 8400prm/1300kHU |
Generator frequency | 30kHZ/400kHZ(optional) |
Radiography table(W*L cm) | 73*200cm, Move freely, the wheels with brake. |

Gravity angle indicator plate·Flexible and precise

Hidden counterweight brake handle·Easy lift
Installation Instructions
1.X-ray room: 20m2 room is recommended. 2. Lead glass size recommended: 40*60cm or 60*80 cm
3. X-ray protection preparation: use barium sulfate powder or lead sheet for the wall of the room, and lead sheet for the ceiling;
4. A fan must be installed on the top to facilitate the discharge of ionized air
5. All surfaces of the x-ray room that people can reach need to be covered with lead sheet or barium sulfate powder. If it is the first floor, the ground does not need to be covered; if it is the top floor and will not polluted the environment, lead protection is not required.