YB-DI-MICRO-10
ISO9001, RoHS, CE
99.0%
Laboratory
80 Gallon
10L/H
up to 2 Liters/Minute(with Pressure Tank)
97%-99% (New RO Membrane)
>99%, When MW>200 Dalton
>99%
10 to 16 M Ω. Cm @ 25 ºC
< 0.1 Us/Cm
< 0.01 Ppb
<30 Ppb
< 0.01 Cfu/Ml
100-240V, 50/60Hz, 48-72W
YBOTECH
Cardboard Box
10L/H
China
8421211000
Product Description

- Applicable to various water quality, the case is with powder coating.
- The filter element adopts open mold injection technology and inside the machine
- Korean quick-insert filter column is easy to replace and maintain.
- The electrical appliance and waterways are completely separated to avoid circuit
- failure caused by moisture and water leakage.
- Online water quality monitoring system, immediately measure the output water quality.
- Automatic RO membrane anti-scaling flushing program, automatic stop when tap water break, automatic water breaking when machine stop, storage tank automatically fill water, automatic stopping when water is full.
- Pipeline has function of upgrading water quality and quantity.
- Equipped with a portable TDS pen, you can test TDS content, conductance, water temperature anytime, anywhere.
- Double electric water intakes, two kinds water quality, ready to use. Pressure online monitoring.
- Open door at front and back, this make filters replacement easy&convenient, the replacement can be easily completed without using any tools. Double water intakes: pure water and deionized water
Product Details


System Filtering Process


Filter element function:
- The first stage: PP cotton (PPF filter element with pore diameter of 5 microns) : preliminary filtration of raw water to remove coarse impurities, colloid, suspended matter, etc.
- Second level: granular activated carbon: efficient adsorption of water odor, heterochromatic, organic matter, some heavy metals, etc.
- Third stage: compressed activated carbon: further remove residual chlorine, organic compounds, heterochromatic, odor, turbidity, etc.
- Level 4: RO reverse osmosis membrane: pore diameter 0.0001 micron, effectively remove viruses, heavy metals and other organic impurities in water.
- Level 5: deionzed purification column ×3, remove most of the heavy metals and trace elements in water, water quality to meet the national laboratory water level 1 water standard.
Product technical parameters
YB-DI-MICRO series deionized water machine | |
Output | 10/20/30 LPH |
Water requirements | Urban tap water, inlet water pressure: 0.15-0.5mpa, water temperature 5-40ºC |
RO water parameter | |
Ion rejection rate | 97%-99% (when using a new RO film) |
Organic matter rejection rate | >99% when MW>200 Dalton |
Particle and bacterial rejection | > 99% |
DI water parameter | |
Resistivity (25ºC) | 10 to 16 m Ω. Cm @ 25 ºC |
Conductivity | < 0.1 us/cm |
Heavy metal ion | < 0.01 ppb |
Total organic carbon (TOC) * | <30 ppb |
bacteria | < 0.01 cfu/ml |
Electrical requirements | 100-240V, 50/60Hz, 48-72W |
The system configuration | Host machine (including 1 set of purification column)+TDS pen + accessories package |
DI water Application field
Preposition unit of ultra pure water system, supply water for biochemical analyzer, electroplating painting, battery charging, chips washing, microbial culture media, chemical&biochemical regent preparation, buffer making, photographic film processing, stability test chamber, meteorological test equipment, hydrogen generator, salt spray test chamber. It can completely replace the traditional single/double/triple distilled water machine.