GA-EDX 9600E
Online
1 Year
Xrf Analyzer
0-100Kg
GONOAVA
Plywood Case
48*40*40cm
China
9024800000
Product Description
GA-EDX 9600E ASTM D4294 XRF Oil Element Composition Analyzer Sulfur and Chlorine Analyzer Tester
Achine advantage
Advanced sulfur-in-oil analyzer
Advanced metals-in-lubricant analyzer
Advanced precision
Advanced reliability and robustness
Advanced software and data management
Overview
The Sulfur in Oil XRF analyzer Complying with ASTM Method 4294
EDX9600E can be used to test and monitor total sulfur concentration in the production of oils and fuels. Compliant, repeatable and accurate elemental analysis is a given but additional value is generated in several parts in your oil production and refinery. EDX9600E complies with international test methods and regulations, such as ISO 13032, ISO 20847, ISO 8754, ASTM D4294, ASTM D4929, JIS K2541-4, IP 336 and IP 496.
The advanced performance of EDX9600E can also be extended to the more applications including
>The wear metals analysis of lubricants to optimize and adapt production processes complied with ASTMD7751/ ASTM D6481/ASTM D5059
>The determination of low levels of chlorine for optimal desalting and corrosion control
>The determination of low levels of S, Ni, V, Ca and Fe to prolong your catalysts' life and improve their performance and the analysis of the sulfur content during fuel production and blending
GA-EDX 9600E ASTM D4294 XRF Oil Element Composition Analyzer Sulfur and Chlorine Analyzer Tester
Features and Benefits
•Rugged elemental analyzer for repeatable and reproducible results in laboratories
production environments
• Intuitive interface is displayed on the large, industrial touchscreen.
• Streamlined software and one-touch measurement start makes it easy for any operator
to get high quality results.
• Training for routine analysis takes only minutes.
• Atmospheric compensation allows analysis without the need for helium or vacuum purge,
minimizing the cost per analysis.
• Automatic carbon/hydrogen (C/H) ratio correction allows the analysis of multiple types of
oils and fuels to be made on a single calibration, simplifying analyzer setup and operation
• X-ray tube and detector are operated in conditions designed to ensure long-term reliability.
• Built for tough environments.
• On-board data storage of up to 100,000 results, including spectra
• Low cost of maintenance
GA-EDX 9600E ASTM D4294 XRF Oil Element Composition Analyzer Sulfur and Chlorine Analyzer Tester
Specifications
High precision results without daily calibration needed
Precision Test report of Sulfur in oil standard samples
Precision Test report of Metals in lubricants standard samples






Achine advantage
Advanced sulfur-in-oil analyzer
Advanced metals-in-lubricant analyzer
Advanced precision
Advanced reliability and robustness
Advanced software and data management
Overview
The Sulfur in Oil XRF analyzer Complying with ASTM Method 4294
EDX9600E can be used to test and monitor total sulfur concentration in the production of oils and fuels. Compliant, repeatable and accurate elemental analysis is a given but additional value is generated in several parts in your oil production and refinery. EDX9600E complies with international test methods and regulations, such as ISO 13032, ISO 20847, ISO 8754, ASTM D4294, ASTM D4929, JIS K2541-4, IP 336 and IP 496.
The advanced performance of EDX9600E can also be extended to the more applications including
>The wear metals analysis of lubricants to optimize and adapt production processes complied with ASTMD7751/ ASTM D6481/ASTM D5059
>The determination of low levels of chlorine for optimal desalting and corrosion control
>The determination of low levels of S, Ni, V, Ca and Fe to prolong your catalysts' life and improve their performance and the analysis of the sulfur content during fuel production and blending
GA-EDX 9600E ASTM D4294 XRF Oil Element Composition Analyzer Sulfur and Chlorine Analyzer Tester
Features and Benefits
•Rugged elemental analyzer for repeatable and reproducible results in laboratories
production environments
• Intuitive interface is displayed on the large, industrial touchscreen.
• Streamlined software and one-touch measurement start makes it easy for any operator
to get high quality results.
• Training for routine analysis takes only minutes.
• Atmospheric compensation allows analysis without the need for helium or vacuum purge,
minimizing the cost per analysis.
• Automatic carbon/hydrogen (C/H) ratio correction allows the analysis of multiple types of
oils and fuels to be made on a single calibration, simplifying analyzer setup and operation
• X-ray tube and detector are operated in conditions designed to ensure long-term reliability.
• Built for tough environments.
• On-board data storage of up to 100,000 results, including spectra
• Low cost of maintenance
GA-EDX 9600E ASTM D4294 XRF Oil Element Composition Analyzer Sulfur and Chlorine Analyzer Tester
Specifications
Detector | High Resolution SDD |
Excitation Source | 50 watts high efficiency Ag/Rh tube 50kv Max,1000Max |
Working temperature | -20 to 50 °C |
Measuring time per sample | 30-130 seconds |
Sample Type | Liquid, powders and solids |
Factory Calibration | Low sulfur: 10ppm- 100ppm High sulfur: 0.01%-5% |
Detection Limit of Sulfur | 3ppm |
Instrument Dimension | 380×372×362mm(L*W*H) |
Weight | 30 kg |
External Connection | USB Port |
Test Report | Excel, PDF |
Spares and Consumables | Sample test Mylar Oil analysis sample cup Sampling spoon Test window film |
High precision results without daily calibration needed
Precision Test report of Sulfur in oil standard samples
Unit: | PPM | Test Time: 100 seconds | ||
Test No. | Calibration Curve | Sulfur-300ppm | Sulfur-1000ppm | Sulfur-10000ppm |
1 | Crude Oil | 297 | 1001 | 10080 |
2 | Crude Oil | 293 | 998 | 9993 |
3 | Crude Oil | 305 | 1017 | 9982 |
4 | Crude Oil | 295 | 1008 | 9918 |
5 | Crude Oil | 291 | 1007 | 9978 |
6 | Crude Oil | 309 | 990 | 10068 |
7 | Crude Oil | 312 | 995 | 9986 |
8 | Crude Oil | 316 | 1015 | 9999 |
9 | Crude Oil | 307 | 1010 | 9918 |
10 | Crude Oil | 313 | 999 | 9928 |
11 | Crude Oil | 302 | 1016 | 9958 |
Certified Value of sulfur(ppm) | 300 | 1000 | 10000 | |
Average Test Result(ppm) | 304 | 1005 | 9983 | |
Standard Deviation Sn | 8.64 | 9.10 | 53.96 | |
RSD | 2.85% | 0.91% | 0.54% |
Precision Test report of Metals in lubricants standard samples
Unit: | PPM | Test Time: 60 seconds | ||||||||
No. | Calibration | Ca | Ti | V | Cr | Mn | Fe | Ni | ||
1 | Lubricant | 318 | 304 | 300 | 307 | 308 | 309 | 308 | ||
2 | Lubricant | 278 | 305 | 294 | 299 | 305 | 303 | 298 | ||
3 | Lubricant | 323 | 302 | 300 | 313 | 303 | 306 | 303 | ||
4 | Lubricant | 279 | 309 | 300 | 304 | 305 | 305 | 299 | ||
5 | Lubricant | 318 | 310 | 300 | 304 | 306 | 305 | 306 | ||
Certified Value | 300 | 300 | 300 | 300 | 300 | 300 | 300 | |||
Average | 303.2 | 306 | 298.8 | 305.4 | 305.4 | 305.6 | 302.8 | |||
Standard Deviation Sn | 22.643 | 3.3912 | 2.6833 | 5.1284 | 1.8166 | 2.1909 | 4.3243 | |||
Error ppm | 3.2 | 6 | -1.2 | 5.4 | 5.4 | 5.6 | 2.8 | |||
RSD | 7.47% | 1.11% | 0.90% | 1.68% | 0.59% | 0.72% | 1.43% | |||
No. | Calibration | Cu | Zn | Mo | Ag | Cd | Pb | |||
1 | Lubricant | 299 | 300 | 315 | 307 | 295 | 297 | |||
2 | Lubricant | 300 | 304 | 306 | 305 | 299 | 297 | |||
3 | Lubricant | 305 | 307 | 308 | 306 | 311 | 297 | |||
4 | Lubricant | 299 | 303 | 306 | 306 | 328 | 298 | |||
5 | Lubricant | 304 | 304 | 305 | 324 | 313 | 294 | |||
Certified Value | 300 | 300 | 300 | 300 | 300 | 300 | ||||
Average | 301.4 | 303.6 | 308 | 309.6 | 309.2 | 296.6 | ||||
Standard Deviation Sn | 2.881 | 2.51 | 4.062 | 8.0808 | 13.008 | 1.5166 | ||||
Error ppm | 1.4 | 3.6 | 8 | 9.6 | 9.2 | -3.4 | ||||
RSD | 0.96% | 0.83% | 1.32% | 2.61% | 4.21% | 0.51% |