Product Description
Medical laboratory equipment fully automatic veterinary hematology analyzer

Description

 
   ItemDescriptions
GeneralThroughput60 samples/hour
PrincipleImpedance for WBC,RBC,PLT;  Cyanide-free method for HGB
Channels2
Aperture diatmeterWBC chamber:100um, RBC/PLT chamber:70um
Parameters20 parameters(including WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%,   
RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR)  
Histograms3 histograms(including WBC,RBC,PLT histograms)
HGB lampLED 540nm
Sample volumeNormal: 10ul, Prediluted: 20ul
SoftwareCalibrationManual & auto calibration
Counting modesVenous, capilary, prediluted
Print Auto print, manual print
LISSupport LIS
LanguageEnglish, Spanish,Italian,Portuguese, French etc
ReagentReagentStandard: Dilent: 20L, Lyse: 500ml
CleanserStandard: Probe cleanser: 50ml
SpecificationsInput8 inch touch screen, keyboard, mouse
OutputInternal printer, support external printer
Printer paper80×35mm
InterfaceRS232 port, USB port, VGA port
Storage100,000 results with histogram
Dimension410mm×435mm×472mm
Blockage clearHigh voltage, high pressure flush
Net weight21 KG
OthersWork temperature18-35°C
Power requirement100~240 AC, 50/60Hz, 150VA
Relative humidity≤85
Atmospheric pressure86.0kPa-106.0kPa 
PerformanceCarry-over rateParameterCV
WBC≤ 2%
RBC≤ 1%
HGB≤ 2%
PLT≤ 2%
BackgroundParameterBackground
WBC≤ 0.2×109/L
RBC≤ 0.02×1012/L
HGB≤ 1 g/L
HCT≤ 0.5%
PLT≤ 10×109/L
PrecisionParameterCVMeasurement range
WBC≤ 2.0%(4.0-15.0×109)/L
RBC≤ 1.9%(3.5-6.0×1012)/L
HGB≤ 1.9%(110.0-180.0g)/L
MCV≤ 0.4%(80.0-110.0)fL
PLT≤ 4.0%(100.0-500.0×109)/L
LinearityParameterMeasurement rangeCV
WBC(0-5.9×109)/L
(6.0-99.9×109)/L
(±0.3×109)/L
±5%
RBC(0-0.99×1012)/L
(1.0-9.99×1012)/L
(±0.05×1012)/L
±5%
HGB(0-99.0g)/L
(100.0-300.0g)/L
(±2.0g)/L
±2%
PLT(0-99×109)/L
100-999×109)/L
(±8.0×109)/L
±10%