TH0530 Through-Core Precision Ammeter

Price $15000.00 Compare
Min Order 1 set
Shipping From Hunan, China
Popularity 426 people viewed
Quantity
+-
 

Tunkia Co., Ltd.

VIP   Audited Supplier 4 years
Profile Certified by SGS/BV
TH0530
Life Time
12 Months Excellent After-Sales Service
Electricity
With LCD Display
Non-Customized
TUNKIA
Package for Exporting
Approx 13 kg
China
Product Description
TH0530 Through-Core Precision Ammeter 

1. Summary

TH0530 through-core precision ammeter is a wide measurement range, high measurement accuracy, small and portable ammeter, using broadband current comparator technology, suitable for measuring broadband high current in the field, or for calibrating AC and DC high current sources and ammeters.
 

2. Features

  1. Available in 1 kA and 2 kA current specifications
  2. Measurement frequency: DC ~ 1 kHz
  3. Accuracy class: 0.005 / class 0.01
  4. Through-core and direct-in measurement
  5. The aperture of Ф 70 mm facilitates penetration of high-current wires
  6. Analog I/V output mode is supported
 

3. Applications

 
  1. Formation of AC and DC current measurement system
  2. DC stable power supply test
  3. Battery charge and discharge test and calibration
  4. Calibrate the multifunction calibrator
  5. Calibrate the AC and DC current source/meter
  6. Clamp meter calibration device traceability
  7. Voltage output reference
  8. Ripple and distortion measurement
 
  1.  
☆ Set up a precision high-current test system
  1. It can be combined with AC and DC test power supply or constant current source to form a precision current test system.
  2. Monitor current in industrial and metering application scenarios, such as battery charge and discharge test systems, regulated power supply test systems, etc
 

☆ Calibrate high current source/meter
  1. TH0530 can be used as a standard ammeter, calibrating AC and DC current sources or ammeters.

4. Characteristics

☆ Through-heart or direct measurement
  1. It has two measurement methods: through-core and straight-in.
  2. The through-core type is suitable for measuring large currents, and is configured with a large aperture to facilitate the connection of high-current test wires.
  3. Small currents can be measured directly to ensure measurement accuracy.
 
☆ Digital and analog signal output
  1. TH0530 supports both digital and analog outputs.
  2. It can directly output secondary analog I/V signals, making data acquisition more flexible and fast.

5. Specifications

5.1 AC direct current measurement

Measurement
manner
Current rangeResolutionFrequency
Hz
Measurement uncertainty (k=2)
(23°C±5°C)
Temperature coefficient (/°C).
(0°C~50°C)
TH0530
-1 kA
TH0530
-2 kA
Class 0.005 Class 
0.01 
Class 0.005 Class
0.01 
(ppm*RD+ppm*RG)[1]
Straight-in2.5 A5 A1 μADC40 + 1080 + 203 + 26 + 4
40≤F≤400150 + 50220 + 80
400<F≤1k350 + 150350 + 150
10 A20 A10 μADC40 + 1080 + 203 + 26 + 4
40≤F≤400150 + 50220 + 80
400<F≤1k350 + 150350 + 150
Piercing250 A500 A100 μADC40 + 1080 + 203 + 26 + 4
40≤F≤400150 + 50220 + 80
400<F≤1k350 + 150350 + 150
1 kA2 kA1 mADC40 + 1080 + 203 + 26 + 4
40≤F≤400150 + 50220 + 80
400<F≤1k350 + 150350 + 150
Note [1]: RD is the reading value and RG is the range value.
  1. Measurement up to 110%* measuring range
  2. Measurement frequency: DC ~ 1 kHz
  3. Range switching: manual/automatic gear shifting
  4. Display digits: 7 digits, decimal
 

5.2 Ripple and distortion measurement

  1. With DC ripple measurement function, frequency range: 1 Hz ~ 10 kHz
  2. Features total harmonic distortion measurement
 

5.3 Secondary voltage output

Measurement
manner
Primary nominal input currentSecondary nominal
Output voltage
Frequency
Hz
Measurement uncertainty (k=2), ppm*RG
TH0530
-1 kA
TH0530
-2 kA
Class 0.005 Class 0.01 
Straight-in2.5 A5 A5 VDC0.0050.01
40≤F≤4000.020.03
400<F≤1k0.050.05
10 A20 A5 VDC0.0050.01
40≤F≤4000.020.03
400<F≤1k0.050.05
Piercing250 A500 A5 VDC0.0050.01
40≤F≤4000.020.03
400<F≤1k0.050.05
1 kA2 kA5 VDC0.0050.01
40≤F≤4000.020.03
400<F≤1k0.0050.01
  1. A maximum output of 5.5 V is possible
 

6. General technical indicators

Power supplyAC:85 V ~ 265 V / 47 Hz ~ 63 Hz
DC:120 V ~ 370 V
Maximum power consumption100 VA
Warm-up time30 minutes
Working environmentTemperature: 0°C ~ 50°C
Humidity: 30% R· H ~ 80%R· H, no condensation.
Others: No electromagnetic field interference.
Storage environmentTemperature: -30°C ~ 70°C
Humidity: 10% R· H ~ 90%R· H, no condensation.
Altitude0 m ~ 3000 m
Communication interfaceRS 232×1
Size175 mm (W) × 330 mm (D) × 275 mm (H)
Display screen5.6-inch color LCD display
QualityApprox 13 kg
 

7. Ordering Information



 
Company Profile
 






Message Type
* Message Content
* Name  
* Phone
* Email
* CAPTCHA