TD1310 Precision AC/DC Standard Meter

Price $74500.00 Compare
Min Order 1 Set
Shipping From Hunan, China
Popularity 327 people viewed
Quantity
+-
 

Tunkia Co., Ltd.

VIP   Audited Supplier 4 years
Profile Certified by SGS/BV
TD1310
Life Time
12 months Excellent After-sales Service
Electricity
With LCD Display
Non-Customized
TUNKIA
Package for Exporting
China
Product Description
TD1310 Precision AC/DC Standard Meter

1. Summary

TD1310 is an AC/DC reference meter, which integrates AC/DC reference voltmeter, AC/DC reference ammeter, small-signal voltmeter, AC/DC electric energy measurement, ripple measurement, harmonic measurement and other functions. It is applicable to calibration of AC/DC current source, AC/DC voltage source, AC/DC power source and assessment of AC/DC electric energy metering device.
 

2. Features

  1. Power/electric energy accuracy:DC class 0.01,AC class 0.02
  2. Voltage/current accuracy:DC class 0.005,AC class 0.01.
  3. DC voltage measurement:10 mV ~ 1150 V.
  4. Direct measurement of DC current:10 μA ~ 120 A.
  5. Support the transducer to expand the current range to 500 A or more
  6. Small signal voltage measurement:0.1 mV ~ 12 V
  7. All electric quantity measurements support full automatic range switching.
  8. Support DC ripple and AC 2nd~63rd harmonic measurement
  9. With the function of comprehensive statistics and analysis of the measured electricity.
  10. Standard electric energy pulse input/output function.
  11. Communication interface:RS232,LAN.
 

3. Applications


Note: When using TD1310 to calibrate electrical  measuring devices or resistors, it is necessary to match the corresponding test power supply.
 
  1. Assessment of electric energy metering device: TD1310 can be used as a high-level electric energy standard meter to assess the verification device of AC electric energy meter, DC electric energy meter, standard electric energy meter, electric vehicle charger tester and other equipment. Its DC+AC measurement mode is also particularly suitable for the verification and calibration of DC electric energy meter verification device with ripple detection function.
  2. Calibration of electrical measurement standard source: Direct verification or calibration of AC/DC voltage source, current source, power source, etc
  3. Calibration of electrical measuring instruments: it can be used as a standard meter to verify or calibrate voltmeter, ammeter, ohmmeter, power meter, power factor meter, frequency meter, phase meter, transmitter, sensor, etc. with a test power supply.

4. Characteristics


 Wide range current measurement
  1. Direct current measurement:Use a pair of quick connect terminals to directly measure 10 μA ~ 120 A(DC); 10 μA~80 A (AC) wide range (AC) current.
  2. Current conversion measurement:It can be equipped with a precision current transducer (TH0540) to expand the current range to 500 A, 1 kA or even more.


 

AC and DC comprehensive measurement
  1. The instrument supports DC, AC and DC+AC measurement modes, and the AC fundamental frequency is 40 Hz~1 kHz.
  2. When measuring voltage or current, the measurement of distortion and ripple can be completed synchronously. The maximum measurement bandwidth of DC ripple reaches 5 kHz. AC supports the measurement of the 2nd~63rd harmonics.
  3. DC+AC measurement function is very suitable for verifying the verification device of electric energy meter with ripple or harmonic output function



 

 Precise small signal voltage measurement
  1. Support small signal voltage measurement of 0.1 mV~12 V (DC) and 2 mV~10.2 V (AC), and can withstand overload signal of 150 Vpk at most.
  2. Application scenario 1: It can be used to calibrate shunt or other I/V transformer with stable large current sources and precision current transducer.
  3. Application Scenario 2: With its own precision voltage measurement function, it can be used to assess the verification equipment for indirect connected DC electrical energy meters.
 

 Power calculation function
 
  1. The instrument has the power calculation function. Users can connect voltage divider, shunt, transformer and other range extenders externally, set the ratio k and offset b, measure the secondary output signal x of the extender through the instrument, and then calculate the actual value of the primary power to be measured according to the function f(x)=kx+b.
 

 DC ripple/AC harmonic measurement
S/NFunction declaration
1With DC ripple measurement function, it can measure ripple content in the range of 1 Hz~5 kHz.
It can display ripple waveform in real time, and measure RMS and P-P of ripple.
2The AC electric quantity measurement also has the oscilloscope function, which can display the waveform of the measured AC signal in real time.
3Calculate THD/T (harmonic vs. full wave) and THD/F (harmonic vs. fundamental wave) harmonic distortion.
4The amplitude (RMS), content (%) and phase of the 2nd~63rd harmonic of each phase voltage or current can be analyzed in real time.
5The spectrum of each harmonic is displayed visually in the form of histogram (fundamental wave is 100%).
 

 Data statistics and analysis function
S/NFunction description
1Data statistics and analysis: Calculate the maximum (Max), minimum (Min), peak to peak (P-P), average (Avg), standard deviation (S.dev), divergence (Div), etc. of the measured electricity.
2Power trend chart test: during the test cycle, draw the curve of power change with time in real time.
3Normal distribution histogram: it displays the distribution of the electricity to be counted during the test period.



 

5. Specifications

5.1 AC and DC Voltage Measuremen

RangeResolutionAccuracy (k=2),(ppm*RD + ppm*RG or V) [1]
DC40 Hz≤F≤200 Hz200 Hz<F≤1 kHzTemperature Coefficient (/°C)
@(15 ~ 30)°C
100mV10 nV50 + 5 μ120 + 80200 + 1003 +0.5 μV
300 mV10 nV30 + 6 μ120 + 80200 + 1003 + 0.6 μV
1 V0.1 μV30 + 2060 + 40120 + 802.5 + 1
3 V0.1 μV30 + 2060 + 40120 + 802.5 + 1
10 V1 μV30 + 2060 + 40120 + 802.5 + 0.5
30 V1 μV30 + 2060 + 40120 + 802.5 + 1
100 V10 μV30 + 2060 + 40120 + 802.5 + 1
300 V10 μV30 + 20120 + 80200 + 1003.5 + 1.5
1000 V0.1 mV30 + 20120 + 80200 + 1003.5 + 1.5
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).
  1. DC measurement range:± (10 mV~1150 V).Manual/automatic range switching
  2. AC measurement range:10 mV~800 V(Valid value).Manual/automatic range switching
  3. Input bandwidth: 50 kHz
  4. Minimum resolution: 10 nV, 8-bit display.
  5. Input impedance:1 MΩ
  6. Input protection:1200 Vpk.
 

5.2 AC and DC Current Measurement

RangeResolutionAccuracy (k=2),(ppm*RD +ppm*RG or nA)
DC40 Hz≤F≤200 Hz200 Hz<F≤1 kHzTemperature Coefficient (/°C)
@ (15 ~ 30)°C
100 μA0.1 nA50+ 5 nA200 + 100250 + 1503 + 0.5 n
300 μA0.1 nA50+ 15 nA200 + 100250 + 1503 + 1.5 n
1 mA1 nA30 + 20200 + 100250 + 1502.5 + 1
3 mA1 nA30 + 2060 + 40100 +502.5 + 1
10 mA10 nA30 + 2060 + 40100 + 502.5 + 1
30 mA10 nA30 + 2060 + 40100 + 502.5 + 1
100 mA0.1 μA30 + 2060 + 40100 + 502.5 + 1
300 mA0.1 μA30 + 2060 + 40100 + 502.5 + 1
1 A1 μA30 + 20200 + 100250 + 1502.5 + 1
3 A1 μA30 + 20200 + 100250 + 1502.5 + 1
10 A10 μA30 + 20200 + 100250 + 1502.5 + 1
30 A10 μA30 + 20200 + 100250 + 1502.5 + 1
100 A0.1 mA30 + 20200 + 100250 + 1502.5 + 1
200 A [2]0.1 mA30 + 20200 + 100250 + 1503.5 + 1.5
500 A [2]0.1 mA30 + 20200 + 100250 + 1503.5 + 1.5
Note [2]: 200 A and 500 A range current are options, and a current transducer such as TH0540 must be selected to expand the range.
  1. DC measurement range:± (10 μA~120 A) .Manual/automatic range switching
The upper limit of DC current can be extended to 240 A or 600 A by connecting the transducer.
  1. AC measurement range:10 μA~80 A(effective value),Manual/automatic range switching
The upper limit of AC current can be extended to 160 A or 400 A (effective value) by connecting the transducer.
  1. Input bandwidth: 50 kHz
  2. Minimum resolution: 0.1 nA, 8-digit decimal number display.

5.3 AC/DC Small Signal Voltage Measurement

RangeResolutionAccuracy (k=2),±(ppm*RD +ppm*RG or μV)
DC40 Hz≤F≤200 Hz200 Hz<F≤1 kHzTemperature Coefficient (/°C)
@ (15 ~ 30)°C
1 mV1 nV40 + 0.2 μ--4 + 0.02 μ
10 mV10 nV40 +0.5 μ120 + 80200 + 1003 + 0.05 μ
100 mV100 nV40 + 1080 + 40100 + 503 + 1.0
1 V1 μV40 + 1080 + 40100 + 502 + 0.5
10 V10 μV40 + 1080 + 40100 + 502 + 0.5
  1. AC measurement range:± (0.1 mV~12 V).Manual/automatic range switching
  2. DC measurement range:2 mV~10.2 V(effective value).Manual/automatic range switching
  3. Input bandwidth: 50 kHz
  4. Minimum resolution: 0.1 nV, 8-digit decimal number display.
  5. Input impedance:> 1 GΩ.
  6. Input protection:150 Vpk,continuity.
 

5.4 AC Frequency/Phase Measurement

TypeRangeAccuracy (k=2)
40Hz ≤ F ≤ 200 Hz200 Hz < F ≤ 1 kHz
Frequency40.000 00 Hz~1000.000 Hz20 ppm20 ppm
Phase0.000 0°~359.999 9°0.02°0.05°
 

5.5 Power/Electric Energy Measurement

DC Power/Electric EnergyMeasuring RangeCombination of DC voltage and DC current
AccuracyUncertainty of DC voltage+uncertainty of DC current (or small signal voltage)
AC Power/Electric EnergyMeasuring RangeCombination of AC voltage and AC current
AccuracyUncertainty of AC voltage+uncertainty of AC current,|PF|=1
Electric Energy PulseStandard Electric Energy Pulse Inputdirectly receive the photoelectric pulse of electric energy of the tested meter
Standard Electric Energy Pulse OutputSignal amplitude is 5V, TTL level
FH=60 kHz × Power value ÷ voltage range value ÷ current range value
FL=60 kHz × Power value ÷ voltage range value ÷ current range value ÷ 10000  
Electric energy error displayAutomatic display, resolution 0.001%
 

6. Ordering Information



 

7. General technical indicators

Power SupplyAC ( 220 ± 22 ) V,( 50 ± 2 ) Hz,Maximum power consumption:100 VA
Temperature PerformanceWorking temperature:15°C~35°C;
Storage temperature:-10°C~50°C
Humidity PerformanceWorking humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C
Storage humidity:(20%~80%) R·H,No condensation
InterfaceRS232,RJ45