High Precision AC Clamp Meter Tester

Price $9858.00 Compare
Min Order 1 Set
Shipping From Beijing, China
Popularity 474 people viewed
Quantity
+-
 

Beijing GFUVE Electronics Co., Ltd.

VIP   Audited Supplier 12 years
Profile Certified by SGS/BV
GF6018A
24h, 7days
3 Years Warranty
Electromagnetic Instrument
Electricity
CE, CCC
0-1100V
0-24A
With LCD Display
Customized
440X450X160mm
12kg
Less Than 85%Rh
50/60Hz
RS232 Interface
LCD Display
2~31
0.05%
GFUVE
Standard Export Carton
CE, ISO9001: 2008, IEC
Beijing, China
Product Description
GF6018A AC CLAMP meter tester with current source and voltage source
Product Description

The equipment is compliance with national verification regulation: JJG124-2005 "Ammeter, voltmeter, power meter and resistance meter verification rules" and related national standards. 0.05 single-phase AC/DC standard source, can verify magnitude 0.2 and the following AC/DC voltmeter, ammeter, frequency meter, resistance meter. The source signals by using DSP and 16-bit high-speed analog-to-digital converters of controllable sine wave, the distorted wave signal source.The equipment comes with RS-232 interface, can be connected with PC to a semi-automatic verification system. It has precise interface and multi-functions including verification, storage and query.The equipment adopts color LCD screen, chart character display is clear, with high precision, stable and reliable, convenient operation and flexible characteristics.

FEATURES

It can be semi-automatic or manual inspection of analog multi meter, clamp type current meter, all kinds of electric meter, (voltmeter, ammeter, frequency meters, resistance meter) basic error
The equipment can be generated with 2-31 harmonic wave, harmonic number, order, amplitude and harmonic phase of fundamental wave can be programmed
Power amplifier working frequency range is 40Hz-1KHZ, has a good linear. Current amplifier is constant current source; voltage amplifier is constant voltage source. Because of light weight, this device is more suitable for field use
With RS-232 interface, this device is controlled by PC through the PC software (optional), can be automatic or manual inspection, and the results are processing and management
Has a non-volatile memory, it can store 500 pieces of the raw data of tested meter, for access and upload

PARAMETERS
Power supplySingle phase AC 220V±10% or 110V±10%, 50/60HzAC
range200mV, 400mV, 2000mV, 4000mV, 
20V, 100V, 200V, 500V, 1000V
Range 20V-1000VAccuracy±(0.03%RD+0.02%FS)
Stability≤0.01% FS /60s
Range 200mV-4000mVAccuracy±(0.3%RD+0.2%FS)
Stability≤0.04% FS /60s
Load capacity20VA
Waveform distortion≤0.3%
Adjustable Range0-120% FS (1000V not included)
Fineness5×10-5
Range2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A
Range 0.2A-20A Accuracy±(0.03%RD+0.02%FS)
Stability≤0.01% FS /60s
Range 2mA-40mA  Accuracy±(0.3%RD+0.2%FS)
Stability≤0.04% FS /60s
Load capacity20VA
Waveform distortion≤0.3%
Adjustable Range0-120% FS
Fineness5×10-5
Range45-65Hz
Fineness0.001Hz
Accuracy0.01Hz
Times2-31
Extent0-20%
Each harmonic phase fineness 0.01·N (N is  harmonic order)
Range200mV, 400mV, 2000mV, 4000mV, 
20V, 40V, 100V, 200V, 400V, 1000V
Load capacity20W
Adjustable range0-110% FS (1000V not included)
Adjustable fineness5×10-5
Stability≤0.01% FS /60s (peak-peak value)
Accuracy±(0.03%RD+0.02%FS)
Ripple wave≤1% FS
Range20uA, 40uA, 200uA, 400uA, 2mA, 4mA, 
20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A
Load capacity20W
Adjustable Range0-110% FS (1000V not included)
Adjustable Fineness5×10-5
Stability≤0.01% FS /60s (peak-peak value)
Accuracy±(0.06%RD+0.04%FS)
Ripple wave≤1% FS
Range0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ, 
0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ
Fineness0.02%
Range 0-2kΩ, 0-20kΩ, 0-200kΩ 
Range 0-2kΩ Adjustable Range0-100%FS
Accuracy±(0.12%RD+0.08%FS)
Range 0-200Ω,0-2MΩAdjustable Range0-100% FS
Accuracy±(0.3%RD+0.2%FS)
Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ
Range 2kΩ-4kΩAdjustable Range5%-100%FS
Accuracy±(0.12%RD+0.08%FS)
Range  200-400Ω, 
2MΩ-4MΩ
Adjustable Range5%-100% FS
Accuracy±(0.3%RD+0.2%FS)
200AAccuracy±0.2%
600AAccuracy±0.5%
1000AAccuracy±1%
Range1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz
Square wave amplitude4.5V
Setting accuracy5×10-5
Dimension (L×W×H) (mm)440x360x160
Weight (kg)12
Operating temperature20°C±10°C
Relative humidity≤85%RH