Technical Data Hemodialysis Machine (W-T 2008)
Non-Customized
CE, ISO13485
Human Blood Processor
Technical Data Hemodialysis Machine (W-T 2008)
380mmx400X1380mm (H*W*D)
88 Kg
0~35oc
-700~+800mmhg,Accurancy +-10mmhg
0 to 2 M Suction Height
-380 Mmhg to +400 Mmhg
1 Mmhg
0~700 Ml/Min
AC220V+-10%,47~63Hz
1.5-6.0 Bar
2 Year
Mermaid
Carton
CE ISO
China
Product Description
Medical Equipment Hemodialysis Kidney Dialysis W-T 2008
Main performance of this device:
1. Self-checking function;
2. Carbonate dialysis;
3. Double needle dialysis;
4. Liquid level detector;
5. Bubble detector;
6. Blood leakage detector;
7. Temperature and electric conductivity monitoring;
8. Arterial pressure, venous pressure and transmembrane pressure monitoring;
9. Rolling blood pump;
10. Heparin pump;
11. Dehydration amount controlled by capacity;
12. Automatic disinfection cleaning program;
13. Stand-by power of blood pump in case of power failure;
14. Information display function of display screen.
Model | W-T 2008 | USE | Dialysis Machine |
Power Supply | AC220V+-10%,47~63Hz | Supply Pressure | 0 To 2M Sunction Height |
Arterial Pressure Working Range | -700~+800mmhg,Accurancy +-10mmhg | Blood Flow Range | 0~700ml/Min |
Tempreture Range | 0~35ºC | Venous Pressure Working Range: | -700~+800mmhg, Accurancy +-10mmhg |
Size | 1380 X 380 X 400 Mm | Weight | 88Kg |
Water Inlet Pressure | 1.5~6.0Bar |
Technical Data
General data:
Dimensions 1380 x 380 x 400 mm (H x W x D);
Weight approx. 88 KG;
Dialysis fluid circuit
Dialysis fluid flow range
Selectable 100-800 mL/min
Dialysis fluid temperature
Selectable 34.0ºC to 40.0ºC
Dialysis fluid conductivity
Range 12 to 16 mS/cm (25ºC)
Accuracy ±0.1 mS/cm
Acid concentration dialysis fluid
Default mixing ratio 1+34 (others possible)
Range 125 to 150 mmol/L
Bicarbonate concentration dialysis fluid
Default mixing ratio 1+27.6 (others possible)
Range -8 to +8 mmol/L bicarbonate