DRX-II-JG
Online Service
1 Year
220V
Hunan Zhenhua
Export Plywood Case
0.5CBM
China
9031809090
Product Description
DRX-II-JG Laser Flash Thermal Diffusivity Thermal Conductivity Instrument
DRX-II-JG laser flash thermal conductivity instrument uses a laser beam to irradiate the sample, and uses an infrared detector (or a rapid temperature couple) to measure the temperature rise on the back of the sample to calculate the thermal diffusivity coefficient of the sample. It has the characteristics of fast and convenient. Its measured thermal diffusion coefficient is 0.001... 10cm2/sec, and the specific heat of the sample can be measured to further calculate the thermal conductivity. Used in the testing of metals and alloys, diamonds, ceramics, graphite and carbon fiber, filled plastics, polymer materials, etc.
The instrument mainly tests thin thermal conductors, solid electrical insulating materials, granular materials, powdery materials, thermal conductivity of coal solid materials, thermal conductive resin, thermal conductive glass fiber and so on. Conforms to GJB 1201.1-91, ASTMD 1461 flash method for measuring thermal diffusivity, ASTM E2585 flash method for measuring thermal diffusivity, GBT 22588 flash method for measuring thermal diffusivity or thermal conductivity.
Principle:
Place the sample in the automatic furnace sampler and heat the furnace to the programmed temperature. At this temperature, the bottom of the sample is illuminated by an energy pulse (xenon lamp), which generates a uniform temperature rise on the top of the sample. A high speed infrared detector is used to measure the temperature rise on the upper surface of the sample.
Main technical parameters:
Model | DRX-II-JG |
Temperature scope | RT ~100ºC RT ~1000ºC RT ~1400ºC RT ~1600ºC for choose |
Sample size | Φ10~25mm Square not less than 10*10mm |
Sample thickness | 0.1mm~ 10mm |
Sample quantity | 1 peice with automatic sampler |
Sample holder | SiC/Al2O3/metal |
Atmosphere | Air |
Thermal diffusion | 0.01 ~1000 mm2/s |
Thermal conductivity | 0.1 W/(m*K) to 2000 W/(m*K) |
Accuracy | ± 4% |
Repeatability | ± 3% |
Pulse source | Nd: YAG Laser, Wavelength: 1064 nm |
Pulse energy | 15J/Pulse adjustable |
Test sample types | Solid block, low viscosity liquid, high viscosity liquid, solid powder (with a fixture), viscous semi-solid, elastic solid, film, etc., can be adapted; |
Testing software | Dedicated analysis and testing software, support WIDOWS10 64-bit operating system |
Main configurations: | 1. Testing host (with infrared thermometer); 2. Heating furnace; 3. Laser controller; 4. Testing software. |