Multichannel Potentiostat Galvanostat 4-Channel Electrochemical Workstation CS310X (Option A)

Price $9800.00 Compare
Min Order 1 set
Shipping From Hubei, China
Popularity 266 people viewed
Quantity
+-
 

Wuhan Corrtest Instruments Corp., Ltd.

VIP   Audited Supplier 7 years
Profile Certified by SGS/BV
CS310X(Option A)
Free
3 Years
Wuhan Corrtest Instruments
4
+/- 1A on Each Channel
+/- 10V
10uhz~1MHz
Energy, Materials, Battery, Corrosion, etc
Support
Corrtest
Standard Carton
4-channel
China
9027899090
Product Description
Corrtest Multichannel potentiostat/ galvanostat/EIS is a versatile instrument offering 8 slots.
Multichannel potentiostat/ galvanostat model CS310X (Option A) is the 4-channel potentiostat. 4 slots are occupied with potentiostat boards. The rest 4 slots are unoccupied for future adding potentiostat board(s). Option A equals that you have 4 sets independent single-channel potentiostat/galvanostat in one chassis. 
EIS module (10μHz~1MHz) is equipped in only ONE channel. Each channel potential control range is ±10V, current control range ±1A, can meet experiment requirement for most people.
Full floating module and electrical isolation design guarantee each channel is totally independent, which ensures accurate data and efficient simultaneous measurements.

                                                                                                 4-channel potentiostat

Application
- Reaction mechanism of Electrosynthesis, electrodeposition (electroplating), anodic oxidation.
- Electrochemical analysis, electro-catalysis, sensor; 
- New energy materials, advanced functional materials, photoelectronic materials;
- Corrosion study of metals in water, concrete and soil etc;
- Fast evaluation of corrosion inhibitor, water stabilizer, coating and cathodic protection efficiency.


There are 4 basic options for CS310X.
Option A: 4-channel with EIS in one channel
Option B: 4-channel, with EIS in all four channels
Option C: 8-channel, with EIS in one channel
Option D: 8-channel, with EIS in all 8 channels

The number of the channel and EIS modules can be cusomized. 
Lower model can be upgraded to higher one by adding more potentiostat boards on current device
EIS module can be upgraded online. 


 
ModelCS310X
OptionsOption AOption BOption COption D
Number of channels 4488
Max. current on each channel±1A ±1A±1A±1A
Potential range on each channel±10V±10V±10V±10V
EIS (10μHz~1MHz)Included on 1 channel onlyIncluded on all 4 channelsIncluded on 1 channel onlyIncluded on all 8 channels
Standard supplyInstrument host CS310X-1 set
CS studio6 software,
Power cable - 1 piece,
Ethernet cable - 1 piece
Electrode cable -8 pcs  
Dummy cell, Manual
Instrument host CS310X-1 set
CS studio6 software,
Power cable - 1 piece,
Ethernet cable - 1 piece
Electrode cable -16 pcs  
Dummy cell, Manual

Specifications 
Specifications
Number of channels with boards: 4Channel insulation resistance: >100MΩ
Communication: EthernetLower-pass filter: covering 8-decade
Potential control range: ±10VConstant current control range: ±1A on each channel
Potential accuracy: 0.1%×full range±1mV Current accuracy: 0.1%×full range
Potential resolution:10μV(>100Hz), 3μV(<10Hz)Current resolution: 1pA
Potential rise time: <1μs(<10mA), <10μs(<2A)Current range: 2nA ~1A, 10 ranges
Reference electrode input impedance: 1012Ω||20pFMaximum current output: ±1A
Compliance voltage: ±21VCurrent increment during scan: 1mA @1A/ms
CV and LSV scan rate: 0.001mV~10000V/sPotential increment during scan: 0.076mV@1V/ms
CA and CC pulse width: 0.0001~65000sDPV and NPV pulse width: 0.0001~1000s
SWV frequency:0.001~100KHzCV minimum potential increment: 0.075mV
AD data acquisition:16bit@1MHz,20bit @1kHzIMP frequency:10μHz~1MHz
DA resolution:16bit, setup time:1μsCurrent and potential range: automatic
Operating System requirements: Windows 10/11Weight: 12.5 Kg  Dimensions: 40*40*14cm   
Electrochemical Impedance Spectroscopy (EIS)
Signal generator
Frequency range: 10μHz~1MHzAC signal amplitude: 1mV~2500mV
Frequency accuracy: 0.005%Signal resolution: 0.1mV RMS
DDS output impedance: 50ΩDC Bias: -10V~+10V
Wave distortion: <1%Waveform: sine wave, triangular wave, square wave
Scan mode: Logarithmic/linear, increase/decrease
Signal analyzer
Maximum integral time:106 cycles or 105sMeasurement delay:0~105S
Minimum integral time:10ms or the longest time of a cycle
DC offset compensation
Potential compensation range: -10V~+10VCurrent compensation range: -1A~+1A
Bandwidth adjustment: automatic and manual, 8-decade frequency range

Techniques
Stable polarization
  • Open Circuit Potential (OCP)
  • Potentiostatic (I-T curve)
  • Galvanostatic
  • Potentiodynamic (Tafel plot)
  • Galvanodynamic (DGP)
Transient Polarization
  • Multi Potential Steps
  • Multi Current Steps
  • Potential Stair-Step (VSTEP)
  • Galvanic Stair-Step (ISTEP)
Chrono Method
  • Chronopotentiometry (CP)
  • Chronoamperametry (CA)
  • Chronocaulometry (CC)
Voltammetry
  • Linear Sweep Voltammetry (LSV)
  • Cylic Voltammetry (CV)
  • Staircase Voltammetry (SCV) #
  • Square Wave Voltammetry (SWV) #
  • Differential Pulse Voltammetry (DPV) #
  • Normal Pulse Voltammetry (NPV)#
  • Differential Normal Pulse Voltammetry (DNPV) #
  • AC Voltammetry (ACV) #
  • 2nd harmonic AC Voltammetry (SHACV)
  • Fourier Transform AC Voltammetry (FTACV)
Battery test
  • Battery Charge and Discharge
  • Galvanostatic Charge and Discharge (GCD)
  • Potentiostatic Charging and Discharging(PCD)
  • Potentiostatic Intermittent Titration Technique (PITT)
  • Galvanostatic Intermittent Titration Technique (GITT)
Electrochemical Impedance Spectroscopy (EIS) (included on ONE channel only)
  • Potentiostatic EIS (Nyquist, Bode)
  • Galvanostatic EIS
  • Potentiostatic EIS (Optional freq.)
  • Galvanostatic EIS(Optional freq.)
  • Mott-Schottky
  • Potentiostatic EIS vs. Time (Single freq.)
  • Galvanostatic EIS vs. Time (Single freq.)
Corrosion Measurements
  • Cyclic polarization curve (CPP)
  • Linear polarization curve (LPR)
  • Electrochemical Potentiokinetic Reactivation (EPR)
  • Electrochemical Noise (EN)
  • Zero resistance Ammeter (ZRA)
Amperometric
  • Differential Pulse Amperometry (DPA)
  • Double Differential Pulse Amperometry (DDPA)
  • Triple Pulse Amperometry (TPA)
  • Integrated Pulse Amperometric Detection (IPAD)
# There is corresponding stripping technique

Simultaneous Measurements
 
You can choose a same electrochemical technique in every channel. As is shown in below picture, we let 4 channels do EIS.



In each channel, the user can also choose different experiments. Set the parameters for experiment in each channel and start simultaneous measurements. As is shown in the below picture, EIS, galvanostatic charge and discharge, CV, and polarization curve tests are conducted at the same time.




Message Type