MRI1212
Insulated
Stranded
MRI Coil Repair
Silver Plated Copper
Nonmagnetic TPU/TPE
FEP
Round Wire
Medical Field
ISO9001, RoHS
Yqf
yqf
Carton
Standard packing or rolling
Shenzhen
8544492100
Product Description
Non Magnetic Non Ferrous MRI compatible24 Wire MRI Cable With 12 Coax 32AWG And 12x 26AWG Signal Wire Diameter 8.5mm


Specification







The 24 Wire MRI Cable With 12 Coax 32AWG And 12x 26AWG Signal Wire is widely used in MRI Machine repair.We also provide other spec MRI cable with different coax numbers and signal wire numbers,such as 12 coax +20 signal wire,16 coax+16 DC wires…..It maybe used in 24-channel head coil for MR imaging
And we are experienced in custom MRI cable for MRI rf coils
Regular MRI compatible Configuration List:
- 10 wire MRI Cable:10x Coax 30AWG (10 Coax RG178)
- 24 wire MRI Cable:12x 32AWG Coax+12×26AWG
- 28 wire MRI Cable :8x 32AWG Coax +20×28AWG
- 32 wire MRI Cable:12x 32AWG Coax +20×28AWG
- 32wire MRI Cable : 16x 30AWG coax RG178+16X26AWG
- 32 wire MRI Cable:12x32AWG Coax +1×21 AWG+19×28AWG
- 48 wire Multi coax cable :16x 30AWG coax RG178+32X 26AWG
- 48 wire Multi coax cable :24x 30AWG coax RG178+24X 26AWG
- 68 wire Multi coax cable: 9x 26AWG coax+9x 32AWG coax+46x28AWG


Specification
Construction And Material
Item | Unit | Specified Value | ||||||||||||||||
32AWG Coaxial (A) | 26AWG Wire(B) | |||||||||||||||||
Number of cores | - | 12 | 12 | |||||||||||||||
Inner Conductor | Material | - | Silver plated Copper wire | |||||||||||||||
Stranding | No./mm | 1/0.20 | 19/0.10 | |||||||||||||||
Diameter(Nom.) | mm | 0.20 | 0.50 | |||||||||||||||
Insulation | Material | - | FEP | FEP | ||||||||||||||
Thick.(Nom.) | mm | 0.20 | 0.15 | |||||||||||||||
Diameter | mm | 0.60±0.02 | 0.80±0.02 | |||||||||||||||
Color | - | Nature | See Fig.1 | |||||||||||||||
Braid Shielding | Material | - | Silver plated Copper wire | - | ||||||||||||||
Strand Dia. (approx.) | mm | 0.08 | ||||||||||||||||
Dia.(approx.) | mm | 0.92 | ||||||||||||||||
Coverage | % | 90% | ||||||||||||||||
Sheath | Material | - | FEP | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 1.18±0.03 | ||||||||||||||||
Color | - | See Fig.1 | ||||||||||||||||
Assembly Diameter(Nom.)1 | mm | 3.3(See Fig.1) | ||||||||||||||||
Assembly Diameter(Nom.)2 | 5.88 | |||||||||||||||||
Tape1 | Material | - | PTFE | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 6.08 | ||||||||||||||||
Braid Shielding | Material | - | Tinned copper wire | |||||||||||||||
Dia.(approx.) | mm | 0.12 | ||||||||||||||||
Coverage | % | 85% | ||||||||||||||||
Tape2 | Material | - | PTFE | |||||||||||||||
Thick.(Nom.) | mm | 0.05 | ||||||||||||||||
Diameter | mm | 6.76 | ||||||||||||||||
Jacket | Material | - | TPU | |||||||||||||||
Thickness (Nom.) | mm | 0.85 | ||||||||||||||||
Diameter | mm | 8.50±0.20 | ||||||||||||||||
Color | - | Gray/Custom |
Technical Data:
• Operating voltage:30V
• Test voltage:AC 1.0kV/1min
• Temperature range:-40°C ~+ 80°C
• Insulation resistance B(20°C):≥ 100MOhm x km
• Characteristic impedance:50±3Ω
• Delay:4.9ns/m
• VSWR:≤1.3(12~300MHz)
• Attenuation typ.(dB/100m):12MHz≤21.7,63MHz≤49.5,
123MHz≤69.1,135MHz≤72.5,300MHz≤108.8 ,350MHz≤155.2,
400MHz≤175.1,450MHz≤190.2,500MHz≤210.1