Specifications: Phaco Emulsification
Phaco Emulsification Frequency: 40KHz±20%;
U/S Handpiece Top End Vibration Accuracy: Max. 120μm±20%;
Stimulating Power Range: 0~100%, soft keyboard input or adjustable stepping, Step: 1%;
Pulse Frequency:1Hz~99Hz, soft keyboard input or adjustable stepping, Step:1Hz; Effective emulsification time:1%~99% (≥5ms);
Soft Keyboard Input or Adjustable Stepping: Step:1%;
display accuracy: ≤±10%;
Burst Time: 5ms~100ms, soft keyboard input or adjustable stepping;
Step:1ms; display accuracy:≤±10%;
Max. Vacuum:66.7kPa (500mmHg);
display accuracy: ≤±15%;
Max. Flow Rate: 40 ml/min;
display accuracy: ≤±20%

High-Frequency Cautery
Rating Frequency:1MHz±20%;
Max. Output Power: 10W±20% (200Ω Load);
Output Adjustable Range: 0~100%, soft keyboard input or adjustable stepping, Step: 1%;
display accuracy: ≤±20%

Vitrectomy
Cutting Frequency: single mode or (20~600) times/min, soft keyboard input or adjustable stepping, step of 5 times/min, display accuracy: ≤ ±15%; Max. Vacuum: 66.7kPa (500mmHg), display accuracy: ≤ ±15%;
Maximum Flux Rate: 40 ml/min, display accuracy: ≤ ±20%

Irrigation/Aspiration
Vacuum Range: 1mmHg~500mmHg;
Flow Rate: 1 ml /min~40 ml /min;
Max. Vacuum: 66.7kPa (500mmHg), display accuracy: ≤ ±15%;
Max. Flow Rate: 40 ml/min, display accuracy: ≤ ±20%;
Reflux Rate: 20 ml/min±20%

 
High-Frequency Cautery
Rating Frequency:
1MHz±20%;
Max. Output Power:
10W±20% (200Ω Load);
Adjustable Range
0~100%,
Vitrectomy
Vacuum Range:
1mmHg~500mmHg
Flow Rate
1 ml /min~40 ml /min;
Max. Flow Rate
40 ml/min, display accuracy
Phaco Emulsification
Phaco Emulsification Frequency:
40KHz±20%
Pulse Frequency
1Hz~99Hz,
display accuracy
: ≤±15%
Effective emulsification time
1%~99% (≥5ms);
Burst Time
5ms~100ms
 
 
 
 
 
 
 
 