Single Pressure System Disposable Pressure Transducer Kit for Edwards

Price $6.00 Compare
Min Order 40 Pieces
Shipping From Guangdong, China
Popularity 207 people viewed
Quantity
+-
 

Shenzhen JCR Medical Technology Limited Company

VIP   Audited Supplier 4 years
Profile Certified by SGS/BV
JIBPT-01-EDW
Sterilization
Disposable
Clinical, Hospital
Plastic
CE, ISO13485, TUV Fcs
Without Power
All
Disposable Pressure Transducer
to Measure The Pressure of Blood
Finished Products
Accurate Reading&Error Controlling
Single /Double/Triple Channel Kit
Hemodialysis Catheter
Te,Imported Chip, Brand Chip-Set
Compatiable Adapter
Multi-Connector&Adapter
with Exhaust Valve
Available
Paper-Poly Pouch/Valve Bag/Blister Box /Custom
JCR
Sterile Blister Box
224*113*45mm
Guangdong, China
90181930
Product Description

Disposable IBP Transducer
1.Consistant accurate working during blood pressure monitoring.
2.100000 grade clean workshop and 100% finial test.
3.Sterile and individual pakage.
4 connectors compatible with most monitors.
5.Clear fluid pathway enables priming duringe monitoring.
6.OEM/ODM are available.
7. ISO CE approval


 
Product Description
Properties
Medical Materials & Accessories
Product Name
Abbott disposable IBP transducer
Brand Name
JCR
Model Number
JIBPT-E-01-UT
Detailed Photos
Main technical para
 
meters and working environment requirements:
Disposable Pressure Transducer
Barometric pressure
70~106Kpa
Relative humidity
10~90% (non-condensing)
Operating pressure range
-50 ~ + 300 mmHg
Sensitivity
5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis
full-scale reading±1.5%
Input impedance
1200Ω ~3200Ω
Zero pressure offset
-20mmHg~+20mmHg
Thermal offset shift
≤±0.3mmHg/°C
Offset drift
After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift
≤±0.1%/°C
Frequency response
Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand
Defibrillation depends on the final connection of the equipment.
Leakage current
Leakage current is dependent on the final connection of the equipment
Overpressure load
-400~+4000mmHg
Shock resistance
Withstand falling from one meter for three times
 
Light sensitivity
When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.
Storage and transport environment requirements
- 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
Contact with the human body
≤168h
Intra-system velocity
Flow
2-5mL/h
Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
Product shelf-life
3 years