Product Description

Specifications    JRBC-50D

  1. Classification: Class I Type CF
  2. Main Frequency: Monopolar:625kHz  Bipolar:625kHz
  3. Working Mode: Intermittent and continuous working, 10s/30s
  4. Output power and peak voltage: 
    1.                       Spec.
      Mode
    Output PowerPeak Voltage
    Power RangeResistance
    Monopolar Pure cutting5~350W500Ω1500V
    Monopolar Blend 15~250W500Ω1500V
    Monopolar Blend 25~200W500Ω1800V
    Monopolar Contact Coagulation5~200W500Ω2500V
    Bipolar Standard Coagulation2~70W150Ω400V
  5. High Frequency Leakage Current
    The neutral electrode is isolated from the earth during the high frequency operation, the High Frequency Leakage Current <150mA.
  6. Output Channels: Monopolar: 1 channel, Bipolar: 1 channel;
  7. Output Setting and Indicating: The output is set by key pressing in different area according to different working mode, it is displayed digitally.
  8.  Output Indicating
    Visual Indicator: Monopolar Cutting - Yellow / Monopolar Coag - Blue / Bipo Coag - Blue
    Audio Indicator: When the power is output, there is an audio indicator, the volume can be adjusted.
  9. Error Indicating
    Visual Indicator: The red indicator indicating the error and danger detected by self-detecting or monitoring (ALARM);
    Audio Indicator: When alarming, there is an pulse audio indicating, this volume cannot be adjusted.
  10. Output Control
    Monopolar: Hand Control, Foot Control;
    Bipolar: Foot Control, Automatic Control.
  11. Main Supply: ~220V, 50Hz
  12. Dimension: 400mm×330mm×160mm
  13. Net Weight: 5.5 kg
  14. Operation Condition:
    Environment Temperature: +5ºC~+40ºC
    Relative Humidity: ≤85%
  15. Transport and Storage Condition:
    Environment Temperature: -40ºC~+50ºC
    Relative Humidity: ≤90%
     