KTS3-500B
24 Hours on Line Services
380V AC
CE, ISO
1 Year
-55ºC~+150ºC
1000×750×700
1900×1962×2200
Within 30min
Within 70min
KOMEG
Strong Plywood Case, Meet International Standards
Standard and customs made
Dongguan, China
90321000
Product Description
Three Zones Climatic Simulation Testing Chamber for Thermal Shock Test
Three zones thermal shock chamber has a hot zone, a cold zone and a test zone. The test zone is used to place samples. The samples stay in the test zone and doesn't move up-down. By switching the damper between the test zone and cold/hot zone to achieve temperature shock test.
High temperature → normal temperature → low temperature
Low temperature → normal temperature → high temperature
Model Number | KTS3-500A | KTS3-500B | KTS3-500D | |
Volume | 500L | |||
Inner Dimensions W×H×D(mm) | 1000×750×700 | |||
External Dimensions W×H×D(mm) | 1900×1962×2200 | |||
Temperature Fluctuation | ≤1ºC | |||
Temperature Range | Low Temperature Chamber | -40ºCto-10ºC | -55ºCto-10ºC | -65ºCto-10ºC |
High Temperature Chamber | +60ºCto+150ºC | +60ºCto+150ºC | +60ºCto+150ºC | |
Max. power | 45KW | 50KW | 70KW | |
Max. Current | 78A | 89A | 125A | |
Power | 380V AC | |||
Damper Switch Time | ≤5sec | |||
Noise | ≤75dB | |||
Temperature Shock Performance | High temperature exposure +150ºC:≥30min Transition time: ≤5min Low temperature exposure -40ºC/ -55ºC/ -65ºC:≥30min | |||
Heating Time | RT~200ºC:about 30min | |||
Cooling Time | A: RT~-60ºC:about 60min B: RT~-70ºC:about 70min D: RT~-75ºC:about 70min |