HP-SCHEM400
Biochemical Apparatus
Biochemistry Analyzer
CE, ISO13485
Middle-aged and Old
340-800nm
Titanium Quartz Alloy, 33UL; Optical Path: 10mm
Thermal Printer, 58mm Width Paper
HP-Schem400
Kinetics, End Point, Two Point, etc.
9kg(Net), 10kg(Gross)
460(L)X380(W)X200(H)mm
RS232 Serial
24 Keys Touch Keyboard
Halogen Lamp 20W/12V, Life Time up to 2000hr
HEPOBIO
Standard Packaging
CE
China
Product Description
Semi-auto Biochemistry Analyzer (HP-SCHEM400)
HP-SCHEM400 Semi-automatic Biochemistry Analyzer is a high performance photometer and can be used in the clinic of hospitals, epidemic prevention stations and health protection stations for biochemical tests.The instrument is keyboard touch screen enables you to rapidly change functions and set parameters, which can be checked on the display without applying difficult commands.Display is a liquid crystal one with alphanumeric and graphic options.Setting and detecting operations are easy and straightforward.





HP-SCHEM400 Semi-automatic Biochemistry Analyzer is a high performance photometer and can be used in the clinic of hospitals, epidemic prevention stations and health protection stations for biochemical tests.The instrument is keyboard touch screen enables you to rapidly change functions and set parameters, which can be checked on the display without applying difficult commands.Display is a liquid crystal one with alphanumeric and graphic options.Setting and detecting operations are easy and straightforward.

Main Advantages: | |
1 | Back spectro photometry |
2 | 8 standard filtersTitanium alloy flow cell with small capacity and minimize cross-contamination |
3 | Large LCD, built-in thermal printer |
4 | Sold in the market for over 16 years, stable performance |
Performance: | |
1 | Back spectro photometry, 8 hard membrane filters, separate signal process circuit for each wavelength, filters select completed electronically, no mechanical movement |
2 | Hermetic solid-state optical system, low contamination of dust and dampness, long life time of hard membrane filters |
3 | Titanium alloy flow cell with small capacity and minimize cross-contamination |
4 | 1000 patients test data and 200 test programs can be saved |
5 | Clock CMOS chip inserted, no noise, high speed, easy maintenance |
6 | 265x65 dot-matrix back light LCD, display character curve simultaneously |
7 | Reasonable software, programmable biochemical test menu, Self-test program |
Technical Specifications: | |
Analysis method | kinetics, end point, two point, etc. |
Light source | halogen lamp 20w/12V, life time up to 2000hr |
Optical system | 340-800nm |
8 standard filter | 340, 380, 405, 492, 510, 546, 578, 630nm wavelength, accuracy:±2nm |
Abs range | -2.500-2.5Abs, resolution: 0.001Abs |
Flow cell | titanium quartz alloy, 33uL; optical path: 10mm |
Temperature | PELTIER elements, 25,30,37ºC±0.1ºC |
Precision | error<1%, drift<5mAbs/hr |
Printer | thermal printer, 58mm width paper |
Keyboard | 24 keys touch keyboard |
Interface | RS232 serial |
Dimensions: | 460(L)x380(W)x200(H)mm |
Weight: | 9Kg(net), 10Kg(gross) |
Working Conditions: | |
Temperature | 0ºC-40ºC |
Humidity | 5-80% |
Power supply | AC220V±15%, 50Hz, 70VA |
Main Test Items List | |
Liver function | ALT AST ALP T-BIL D-BIL TP ALB TTT CHE Nh3 γ-GT |
Renal function: | BUN CRE UA CO2-CP |
Lipids | T-CHO TG HDL-C LDL-C |
Cardiac function | CK CK-MB LDH HDBH |
Others: | GLU K Na Cl Ca |